APS-7000 Programming Manual
146
-201 Invalid while
in local
Indicates that a command is not executable while
the device is in local due to a hard local control
(see IEEE 488.2, 5.6.1.5); for example, a device with
a rotary switch receives a message which would
change the switches state, but the device is in local
so the message cannot be executed.
Indicates that a legal password-protected program
command or query could not be executed because
the command was disabled.
Indicates that a GET, *TRG, or triggering signal
was received and recognized by the device but
was ignored because of device timing
considerations; for example, the device was not
ready to respond. Note: a DT0 device always
ignores GET and treats *TRG as a Command Error.
Indicates that a request for a measurement
initiation was ignored as another measurement
was already in progress.
Indicates that a program data element related error
occurred. This error message should be used when
the device cannot detect the more specific errors
described for errors -221 through -229.
Indicates that a legal program data element was
parsed but could not be executed due to the
current device state (see IEEE 488.2, 6.4.5.3 and
11.5.1.1.5.).
Indicates that a legal program data element was
parsed but could not be executed because the
interpreted value was outside the legal range as
defined by the device (see IEEE 488.2, 11.5.1.1.5.).