Calibration
WTX110 A4774-1.0 HBM: public 173
Information
The Clr key can be used to clear the value and enter a new one.
Information
To save the zero balance before calibrating the weighing range, you must exit
Setup at this point. Answer the Save Parameters prompt with Y. Then you
can call group 2 again to resume the calibration, skipping the zero calibration.
Menu Key Function/Secondary selection
Calibrate Load? Y Calibrate weighing range
Y Calibrate range
N Continue to step Load(mV/V)
Notice: You are recommended to use as large a
calibration weight as possible.
Calibr.Weight 999999 Suggestion for calibration weight (=weighing
range)
Clear the suggested weight value and enter any
partial load value
Mount load and confirm; continue with calibration
Calibrating... Measurement of the load signal
Load: 9999999 Display the 10 times higher weight value
resolution (as a check)
Load (mV/V): 1.40000 Display the standardized signal - see Calculate
Span
Notice: The signal can be calculated in millivolts
from the displayed mV/V value by multiplying it
by the Excitation value (5V or 10V).