Calibration
172 A4774-1.0 HBM: public WTX110
Menu Key Function/Secondary selection
Geo value 99 Enter the geo value. Values between 0 and 31
can be set, according to the geo value table. The
geo value of the calibration location is set. If
differing, the geo value of the installation location
can be entered subsequently.
Calibrate Zero? Y Zero balance of the scale: Unload scale
Y Calibrate zero point
N Continue to step Zero(mV/V)
Calibrating... Measure the zero point of the scale
Zero: 0.00 Display the 10 times higher weight value
resolution (as a check)
Zero(mV/V): 0.40000 Display the standardized signal - see Calculate
Span.
Notice: The signal can be calculated in millivolts
from the displayed mV/V value by multiplying it by
the Excitation value (5V or 10V).