Calibration
180 A4774-1.0 HBM: public WTX110
Function/Secondary selectionKeyMenu
Overload 9999 Overload display suppression (in D) can be set
here. If the scale exceeds the value (maximum
load + overload), the display is hidden.
Example: Overload 9
On multiple-range scales, 1d corresponds to the
digit of the lowest weighing range.
Incline Switch Off OFF No incline switch connected
In 0…In7 Discrete input
Sensor Connection to an incline switch for
correction or switch off
For more information see chapter 11.9
NTEP approval? N Switch to the metrological properties according to
American requirements (National Type
Evaluation Program) as per Handbook 44.
N Off (default)
Y On
Must be set to Y(es) for applications in the USA,
Canada and Australia!
Underload 20d? Y Hide weight indicator if gross weight values less
than -20d (below zero)
N Off
Y On
For applications in accordance with OIML R76,
Underload 20d must be set to Y
With Taring? Y Select whether tare functions are disabled or
enabled.
N Tare functions disabled
Y Tare functions enabled (default)
For scales that only operate in Gross mode, the
tare functions can be disabled here so as to skip
cost-intensive checking of the tared scale during
legal verification.