Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 2 - 1
Service Manual (MAN272 REV 4)
2
Chapter 2
Troubleshooting Procedures
Getting Started
WARNING:
Only facility-authorized personnel should troubleshoot the bed.
Troubleshooting by unauthorized personnel could result in personal
injury or equipment damage.
Begin each procedure in this chapter with step 1. Follow the sequence outlined
(each step assumes the previous step has been completed). In each step, the
normal operation of the product can be confirmed by answering Yes or No to
the statement. Your response will lead to another step in the procedure, a repair
analysis procedure (RAP), or a component replacement. If more than one
component is listed, replace them in the given order.
To begin gathering information about the problem, start with Initial Actions.
To isolate or identify a problem and to verify the repair after completing each
corrective action (replacing or adjusting a part, seating a connector, etc.),
perform the Function Checks.
To verify the repair, perform the Final Actions after the Function Checks.
If troubleshooting procedures do not isolate the problem, contact Hill-Rom
Technical Support.