Function Checks
Chapter 2: Troubleshooting Procedures
Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 2 - 9
Service Manual (MAN272 REV 4)
2
Yes No
Go to RAP 2.7.
Air System Functional Check
1. Press the seat inflate button, and activate the bed air surface. The seat
section mattress inflates.
Yes No
Go to RAP 2.12.
2. Press and hold the seat deflate button, and activate the bed air surface. The
seat section mattress deflates.
Yes No
Go to RAP 2.12.
3. Press the back inflate button, and activate the bed air surface. The lumbar
section inflates.
Yes No
Go to RAP 2.12.
4. Press the back deflate button to activate the bed air surface. The lumbar
section deflates.
Yes No
Go to RAP 2.12.
Other Bed Functional Checks
1. The lockout switch locks out the designated inboard/outboard controls
when placed in the “OFF” position (locked), and enables the functions
when placed in the “ON” position (unlocked). The lockout functions work
properly.
Yes No
Replace the faulty rocker switch (refer to procedure 4.26). If this
solves the problem, go to “Final Actions” on page 2-10.
Otherwise, contact Hill-Rom Technical Support.