EasyManua.ls Logo

Hill-Rom Affinity Three - Night Light Does Not Illuminate

Hill-Rom Affinity Three
308 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2.13 Night Light Does Not Illuminate
Chapter 2: Troubleshooting Procedures
Page 2 - 34 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 4)
2.13 Night Light Does Not Illuminate
1. The night light comes on when the ambient light dims.
Yes No
Replace the lamp bulb or fixture if damaged (refer to procedure
4.13). If this solves the problem, go to “Final Actions” on page
2-10. Otherwise, continue to step 2.
2. Verify that the photocell functions. The night light comes on when the
ambient light dims.
Yes No
Replace the photocell. If this solves the problem, go to “Final
Actions” on page 2-10. Otherwise, continue to step 3.
3. The system functions normally.
Yes No
Verify that the night light lamp cable is plugged into the logic
control P.C. board connector, P1. If this solves the problem, go
to “Final Actions” on page 2-10. Otherwise, continue to step 4.
4. The system functions normally.
Yes No
Replace the logic control P.C. board (refer to procedure 4.23).
If this solves the problem, go to “Final Actions” on page 2-10.
Otherwise, contact Hill-Rom Technical Support.
5. Go to “Final Actions” on page 2-10.

Table of Contents