2.17 CPR Release Malfunction
Chapter 2: Troubleshooting Procedures
Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 2 - 39
Service Manual (MAN272 REV 4)
2
2.17 CPR Release Malfunction
1. Pull and hold the CPR release handle. The head section goes into the full
flat position. (When unoccupied, the head section may require slight
downward hand pressure to rotate completely into the full flat position.)
Yes No
Check the adjustment of the CPR function (refer to procedure
4.8). If this solves the problem, go to “Final Actions” on page
2-10. Otherwise, go to step 2.
2. Visually inspect the head drive motor assembly for damage or missing
parts. The head drive motor assembly is in good working condition.
Yes No
Replace the head drive motor assembly (refer to procedure 4.5).
If this solves the problem, go to “Final Actions” on page 2-10.
Otherwise, go to step 3.
3. Another part of the bed interferes with the CPR release mechanism.
Yes No
If the head section still does not release, contact Hill-Rom
Technical Support.
4. Take appropriate actions to eliminate the interference.
This solves the problem.
Yes No
Contact Hill-Rom Technical Support.
5. Go to “Final Actions” on page 2-10.