2.10 Hilow Up Malfunction (When The Foot Section Is Lowered)
The hilow function (or motor) fails to automatically raise the bed from the low
position when the foot section is lowered to its lowest point.
1. Activate the hilow function from the siderails and pendant control. The
hilow function works properly.
Yes No
Go to “Hilow Up Malfunction” on page 2-12 or “Hilow Down
Malfunction” on page 2-14. If this solves the problem, go to
“Final Actions” on page 2-10. Otherwise, continue to step 2.
2. Put the bed into the low position. Lower the foot section to the low
position. The bed automatically rises.
Yes No
Check the hilow and foot limit switches. Refer to “Hilow Limit
Switch” on page 4-11 or “Foot Limit Switch” on page 4-22. If
this solves the problem, go to “Final Actions” on page 2-10.
Otherwise, contact Hill-Rom Technical Support.
3. Go to “Final Actions” on page 2-10.