5.3 Measurement specifications
148
D/A Output Specifications
(included with PW3335-02 and PW3335-04, specify at time of order)
Number of output channels 7
Configuration 16-bit D/A converter (polarity + 15 bits)
Output parameters Output parameters for all channels
Select from: Voltage, current, active power, apparent power, reactive power, power fac-
tor, phase angle, frequency, total harmonic voltage distortion, total harmonic current
distortion, voltage crest factor, current crest factor, time average current, time average
active power, voltage ripple rate, current ripple rate, current integration, active power in-
tegration, maximum current ratio
The rectifier can be selected.
Harmonic-order output is not supported.
Output accuracy f.s.: Relative to the output voltage rated value for each output parameter
Level output: (Output parameter measurement accuracy) + (±0.2%f.s.)
High-speed level output: (Output parameter measurement accuracy) + (±0.2%f.s.)
Waveform output: (Output parameter measurement accuracy) + (±1.0%f.s.)
Instantaneous voltage, instantaneous current: RMS value level
Instantaneous power: Average value level
Output frequency band Waveform output, high-speed level output
At DC or 10 Hz to 30 kHz, accuracy is as defined above.
Conditions of guaranteed
accuracy
Temperature and humidity range : 23°C±C (73°F±9°F), 80% RH or less
Warm-up time: 30 minutes, after zero-adjustment
Output voltage The output level, output speed, and waveform output can be selected.
Level output 2 Vf.s.(STD.2):±2 V DC at full scale (f.s.), linked to display updates
Level output 5 Vf.s.(STD.5):±5 V DC at full scale (f.s.), linked to display updates
High-speed level output 2 Vf.s.(FAST.2):
±2 V DC at full scale (f.s.), linked to synchronization interval
High-speed level output 5 Vf.s.(FAST.5)
±5 V DC at full scale (f.s.), linked to synchronization interval
Waveform output 1 Vf.s.(FAST):1 V f.s. at full scale, linked to sampling
Available selections vary with the output parameter.
Voltage, current, active power
Select from: STD.2/ STD.5/ FAST.2/ FAST.5/ FAST
However, specifying FAST.2 or FAST.5 forcibly sets the rectifier to AC+DC.
Apparent power, reactive power, power factor, phase angle, total harmonic voltage
distortion, total harmonic current distortion, voltage ripple rate, current ripple rate,
voltage crest factor, current crest factor, time average current, time average active
power, maximum current ratio
Select from: STD.2/ STD.5
Frequency, current integration, active power integration
fixed to STD.5
Full scale (f.s.) for the above parameters is as follows:
Voltage, current, active power, apparent power, reactive power, time average current,
time average active power:
measurement range
For time average current or time average active power, setting
to TOTAL range causes current measurement range to be set
to 20 A, and power measurement to voltage range × 20 A.
Power factor: ±0.0000 (±1.0000: DC0 V)
Phase angle: ±180.00°
Voltage ripple rate, current ripple rate, total harmonic voltage distortion, total harmon-
ic current distortion: 100.00%
Voltage crest factor, current crest factor:10.000
Maximum current ratio:10.000
Frequency: Selected from the following:
0.5 Hz/ 5 Hz/ 50 Hz/ 500 Hz/ 5 kHz/ 50 kHz/ 500 kHz
Current integration, active power integration:
Selected from the following:
5 m/ 50 m/ 500 m/ 5/ 50/ 500/ 5 k/ 50 k/ 500 k/ 5 M/ 50 M/
500 M/ 5000 M [Ah/Wh]
Error output: Level output of approx. ±9 V DC is generated when the output parameter
display is [o.r] or when the PEAK OVER U or PEAK OVER I lamp lights up.
Output of 0 V is generated for [- - - - -] (invalid data display).