EasyManua.ls Logo

Hioki PW3335 - Page 167

Hioki PW3335
192 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5.5 Calculation Formulas Specifications
161
(3) Power calculation formulas
Item (RECTIFIER) Calculation formulas
Active power
(AC+DC, AC+DC Umn)
Active power simple average value
(DC)
Active power AC component (AC)
Fundamental wave active power
(FND)
P
1
from harmonic calculation formula
Apparent power
(AC+DC, AC, AC+DC Umn)
Calculated values from the respective rectifiers are used for
U and I.
(For AC+DC Umn, the AC+DC value is used for current.)
Fundamental wave apparent power
(FND)
S
1
from harmonic calculation formula
Reactive power
(AC+DC, AC, AC+DC Umn)
Calculated values from the respective rectifiers are used for P and S.
Fundamental wave reactive power
(FND)
Q
1
from harmonic calculation formula
Power factor
(AC+DC, AC, AC+DC Umn)
Calculated values from the respective rectifiers are used for P and S.
Fundamental wave power factor
(FND)
λ
1
from harmonic calculation formula
Phase angle (AC)
Calculated values from the respective rectifiers are used for
λ
.
Fundamental wave voltage current
phase difference (FND)
M: Number of samples between synchronized timing points; s: Number of sample points;
si: Polarity symbol, acquired based on voltage waveform and current waveform lead and lag
Symbol [None]:Current lags voltage (LAG).
Symbol [-]: Current leads voltage (LEAD).
1M
0=s
ss
)×(
M
1
= iuP   
DCDCDC
×= IUP
DCAC
= PPP
IUS ×=
22
= PSQ
si
S
P
λ
si
=
)° 90± °0(  
λφ
1
si
cos=
When P
AC
0
)° 180± °90±(  
  
λφ
1
si
cos180=
When P
AC
< 0
from harmonic calculation formula
1
   φ

Table of Contents