3.2 Configuring Settings
55
• When the frequency of the input to the set synchronization source is less than 5 Hz, the data update (dis-
play update) timing will vary with the frequency of the input to the synchronization source.
Example: If the frequency of the input to the synchronization source is 0.8 Hz, the data (display) will be
updated every 1/0.8 = 1.25 sec.
• If the SYNC lamp (ERROR) lights up when the timeout has been set to a value other than 0.1 sec., the
display will be updated once every time the timeout set time elapses.
• The timeout setting cannot be changed while integration is being performed or during display hold or max-
imum value/minimum value hold operation.
• The instrument is equipped with the high-pass filters to avoid influences from DC components of input sig-
nals during detecting cycles. Because the time constants of the high-pass filters are linked to the timeout
settings, if the range is changed, or an input voltage or current including a DC component rapidly changes,
it will take a time until the measured value stabilizes.
Please wait and read the measured value after the following time. The required time varies depending on
the timeout setting.
• When the timeout is set to 0.1 sec.: approximately 0.6 sec.
• When the timeout is set to 1 sec.: approximately 10 sec.
• When the timeout is set to 10 sec.: approximately 40 sec.
5 Press to exit the settings.
The instrument will return to the normal display (Mea-
surement screen).