When the frequency of the input to the set synchronization source is less than 5 Hz, the data update (dis-
play update) timing will vary with the frequency of the input to the synchronization source.
Example: If the frequency of the input to the synchronization source is 0.8 Hz, the data (display) will be
updated every 1/0.8 = 1.25 sec.
If the SYNC lamp (ERROR) lights up when the timeout has been set to a value other than 0.1 sec., the
display will be updated once every time the timeout set time elapses.
The timeout setting cannot be changed while integration is being performed or during display hold or max-
imum value/minimum value hold operation.
The instrument is equipped with the high-pass filters to avoid influences from DC components of input sig-
nals during detecting cycles. Because the time constants of the high-pass filters are linked to the timeout
settings, if the range is changed, or an input voltage or current including a DC component rapidly changes,
it will take a time until the measured value stabilizes.
Please wait and read the measured value after the following time. The required time varies depending on
the timeout setting.
When the timeout is set to 0.1 sec.: approximately 0.6 sec.
When the timeout is set to 1 sec.: approximately 10 sec.
When the timeout is set to 10 sec.: approximately 40 sec.
5 Press to exit the settings.
The instrument will return to the normal display (Mea-
surement screen).