EasyManua.ls Logo

Kaye Validator AVS - Page 36

Kaye Validator AVS
170 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 2. The Validator AVS Hardware
Kaye Validator AVS User’s Manual 25
2.3.1a Connecting a Thermocouple
Connect types J, K and T thermocouples to the connectors as shown in Figure 10 below.
Cut through the outer insulation to separate the red and blue wires, and then strip back each
wire approximately ½ inch in order to make the connection with the connector.
+ + + + + +
1 2 3 4 5 6
Figure 10: Thermocouple Connections
Always connect the positive (+) lead to the positive (+) connector and the negative (-) lead
to the negative (-) connector. The negative thermocouple lead is normally red, in
conformance with ANSI standards.
Moist Heat Environments
When validating moist heat processes, a sealed PTFE tip thermocouple should be used
with a drip cut (shown in Figure 11 below) on the outer insulation close to the SIM module
to reduce the possibility of drawing moisture into the SIM. To add a drip cut, remove 4
inches (10 cm) of the outer insulation from each thermocouple at a point where natural
drainage can take place without water reaching the terminal screws (see Figure 11 below).
Figure 11: Drip Cut

Table of Contents

Related product manuals