Chapter 6. Defining Study Setups
Kaye Validator AVS User’s Manual 68
For voltage, you can enter up to 3 decimal values in the input and output fields. The default
values are: Input Min = 0.00 V, Input Max = 10.00 V, Output Min = 0.00 Output Max = 10.00.
The input voltage range can be -1 to 12 V.
For current, you can enter up to 3 decimal values in the input and output fields. The default
values are: Input Min = 4.00 mA, Input Max = 20.00 mA, Output Min = 0.00, Output
Max = 10.00. The range of input current is from 1 up to 25 mA.
For pressure:
You can enter up to 3 decimal values in the voltage input and output fields. The default
values are: Input Min = 0.00 V, Input Max = 10.00 V, Output Min = 0.00 Output Max =
10.00. The input voltage range can be -1 to 12 V.
You can enter up to 3 decimal values in the current input and output fields. The default
values are: Input Min = 4.00 mA, Input Max = 20.00 mA, Output Min = 0.00, Output Max =
10.00. The range of input current is from 1 up to 25 mA.
In the last step the sensor input is marked with a Description sensor label number, and to
enable/disable auto numbering. Auto numbering is enabled only if more than one sensor is
selected. After the sensor configuration pressing the Group Sensors tab to the right advances
to the next screen.