EasyManuals Logo

Keithley 2701 User Manual

Keithley 2701
506 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #392 background imageLoading...
Page #392 background image
15-10 SCPI Reference Tables Model 2701 User’s Manual
[SENSe[1]]
:VOLTage:AC Path to configure AC voltage. Sec 3
:APERture <n> [, <clist>] Set integration rate in seconds
(3.333333e-5 to 1).
(Note 2) Sec 4
:APERture? [<clist>] Query aperture integration rate.
:NPLCycles <n> [, <clist>] Set integration rate in line cycles (60Hz;
0.002 to 60, 50Hz; 0.002 to 50).
5.0 Sec 4
:NPLCycles? [<clist>] Query line cycle integration rate.
:RANGe Path to set measurement range: Sec 4
[:UPPer] <n> [, <clist>] Select range (0 to 757.5). 750
[:UPPer]? [<clist>] Query range.
:AUTO <b> [, <clist>] Enable or disable auto range. ON
:AUTO? [<clist>] Query state of auto range.
:DIGits <n> [, <clist>] Specify measurement resolution (4 to 7). 6 Sec 4
:DIGits? [<clist>] Query resolution.
:REFerence <n> [, <clist>] Specify reference (-757.5 to 757.5). 0 Sec 5
:STATe <b> [, <clist>] Enable or disable reference. OFF
:STATe? [<clist>] Query state of reference.
:ACQuire [, <clist>] Use input signal as reference.
:REFerence? [<clist>] Query reference value.
:AVERage Path to configure and control filter: Sec 4
:TCONtrol <name> Select filter type: (MOVing or REPeat). (Note 3)
:TCONtrol? Query filter type.
:WINDow <NRf> Set filter window in % of range
(0 to 10).
0.1
:WINDow? Query filter window.
:COUNt <n> [, <clist>] Specify filter count (1 to 100). 10
:COUNt? [<clist>] Query filter count.
[:STATe] <b> [, <clist>] Enable or disable filter. (Note 4)
[:STATe]? [<clist>] Query state of digital filter.
:DETector Path to set bandwidth: Sec 4
:BANDwidth <NRf>
[, <clist>]
Set AC detector bandwidth in Hertz
(3 to 3e5).
30
:BANDwidth? [<clist>] Query bandwidth.
Table 15-5 (continued)
SENSe command summary
Command Description
Default
parameter Ref SCPI

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
Max Voltage300 V
Max Current3 A
Frequency Range3 Hz to 500 kHz
Frequency Accuracy0.01%
Weight3.6 kg
Measurement FunctionsFrequency, Temperature
InterfaceIEEE-488 (GPIB), RS-232
DC Voltage Range300 V
DC Current Range3 A
AC Voltage Range300V
AC Current Range1 μA to 3 A
Resistance Accuracy0.01%
Channels10
Dimensions213mm x 88mm x 370mm