15-22 SCPI Reference Tables Model 2701 User’s Manual
SYSTem
:CARDX
:SNOpen? Query whether card is “single no-open” type
(i.e., 7711): 1 = yes, 0 = no.
:BANKs? For “single no-open” card, query number
of banks. If not “single no-open” type,
error -221 (settings conflict) results.
:SWOpen? Query whether card is “single with-open”
type (i.e., 7711): 1 = yes, 0 = no.
:BANKs? For “single with-open” card, query number
of banks.
:CSOhms? Query whether card supports common-side
4-wire ohms (i.e., 7701).
:TIME <hr, min,sec> Set system time using 24-hour format. Sec 6
:TIME? Query system time.
:DATE <yr, mo, day> Set system date (yr = 20xx). Sec 6
:DATE? Query system date.
:TSTamp Path to set timestamp: Sec 6
:TYPE <name> Select timestamp type (RELative or
RTCLock).
REL
:TYPE? Query timestamp type that will be used for
the next buffer storage.
:RELative Path to reset relative timestamp:
:RESet Reset relative timestamp to 0.
:RNUMber Path to reset reading number: Sec 14
:RESet Reset reading number. Next reading will
be #1.
:ERRor? Query Error Queue. (Note 3) Sec 11 ✓
:CLEar Clears messages from the Error Queue. Sec 11
:VERSion? Query rev level of SCPI standard. Sec 14 ✓
:LOCal Take 2701 out of remote and restore operation
of front panel controls.
:REMote Place 2701 in remote.
:RWLock Lockout front panel controls.
:RESet Immediately reboots the 2701.
Table 15-7 (continued)
SYSTem command summary
Command Description
Default
parameter Ref SCPI