Model 2701 User’s Manual Getting Started 1-23
Saving a power-on setup
1. Configure Model 2701 for the desired measurement application.
2. Press SHIFT and then SAVE to access the save setup menu.
3. Press the key to display the present power-on (PWR-ON) setup; FACT, *RST,
SAV0, SAV1, SAV2, SAV3, or SAV4.
4. Press to place the cursor on the present power-on setup.
5. Use the
or key to display the desired setup and press ENTER. The instrument
returns to the normal measurement state.
Restoring a setup
1. Press SHIFT and then SETUP to access the restore setup menu.
2. Press to place the cursor on the present RESTORE setup (FACT, *RST, SAV0,
SAV1, SAV2, SAV3, or SAV4).
3. Use the or key to display the desired setup and press ENTER. The instrument
returns to the normal measurement state.
NOTE If the settings for a user setup or power-on setup do not match the switching
module types presently installed in the Model 2701, error +520 (Saved setup
scancard mismatch) occurs when the setup is recalled. The scan list will reset to
the factory defaults and all channels will open. However, the saved setup is still
retained in memory and can be restored when the matching switching module is
later installed.