Performance Tests 1
Keysight 53210A/53220A/53230A Assembly Level Service Guide 41
9 Press the Channel 1 input key.
– The following softkeys are displayed:
10 Press and set:
– Coupling: DC
– Input Impedance: 50 Ω
– Range: 5 V
– Level Setup / Auto Level: Off
– Level Setup / Input Level: 0.000 V
–BW Limit: Off
–Probe: None
11 Press the Math key.
–The Math softkeys are displayed:
12 Press to set:
–All Math: On