Triggers 11
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 101
 
Using Autoscale to
Set Up Edge
Triggers
The easiest way to set up an Edge trigger on a waveform is to use Autoscale. 
Simply press the [Auto Scale] key and the oscilloscope will attempt to trigger on 
the waveform using a simple Edge trigger type. See "Use Autoscale" on page 24.
Alternating edge mode is useful when you want to trigger on both edges of a clock (for 
example, DDR signals).
Either edge mode is useful when you want to trigger on any activity of a selected source.
All modes operate up to the bandwidth of the oscilloscope except Either edge mode, which 
has a limitation. Either edge mode will trigger on Constant Wave signals up to 100 MHz, but 
can trigger on isolated pulses down to 1/(2*oscilloscope's bandwidth).
MegaZoom Technology Simplifies Triggering
With the built-in MegaZoom technology, you can simply Autoscale the waveforms, then stop 
the oscilloscope to capture a waveform. You can then pan and zoom through the data using 
the Horizontal and Vertical knobs to find a stable trigger point. Autoscale often produces a 
triggered display.