188 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
16 Mask Testing
:MTES:SCAL:BIND 0
:MTES:SCAL:X1 -400.000E-06
:MTES:SCAL:XDEL +800.000E-06
:MTES:SCAL:Y1 +359.000E-03
:MTES:SCAL:Y2 +2.35900E+00
Building a Mask File
The following mask uses all eight mask regions. The most difficult part of creating
a mask file is normalizing the X and Y values from the time and voltage values.
This example shows an easy way to convert voltage and time to normalized X and
Y values in the mask file.
The following mask file produced the mask shown above:
MASK_FILE_548XX
"All Regions"
/* Region Number */ 1
/* Number of vertices */ 4
-12.50, MAX
-10.00, 1.750
10.00, 1.750
12.50, MAX
/* Region Number */ 2
1
8
23
4
5
6
7