Triggers 11
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 107
 
5 Press the Rising Edge or Falling Edge softkey to toggle between edge types.
6 Press the Level Select softkey to select High; then, turn the Trigger Level knob to 
adjust the high level.
7 Press the Level Select softkey to select Low; then, turn the Trigger Level knob to 
adjust the low level.
You can also push the Trigger Level knob to toggle between High and Low 
selection.
8 Press the Qualifier softkey to toggle between "greater than" or "less than".
9 Press the Time softkey, and turn the Entry knob to select the time.
Setup and Hold Trigger
On DSOX1200-Series oscilloscopes, the Setup and Hold trigger looks for setup 
and hold violations.
One oscilloscope channel probes the clock signal and another channel probes the 
data signal.
To trigger on setup & hold violations:
1 Press the [Trigger] key.
2 In the Trigger menu, press the Trigger Type softkey.
3 Press the Trigger Type softkey; then, turn the Entry knob to select Setup and Hold.
4 Press the Clock softkey; then, turn the Entry knob to select the input channel 
with the clock signal.
5 Set the appropriate trigger tevel for the clock signal using the Trigger Level 
knob.
Data
Clock
(rising
edge)
Setup time Hold time