EasyManuals Logo

Keysight DSOX1202A User Manual

Keysight DSOX1202A
330 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #77 background imageLoading...
Page #77 background image
77
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes
User's Guide
7 Reference Waveforms
To save a waveform to a reference waveform location / 77
To display a reference waveform / 78
To scale and position reference waveforms / 79
To adjust reference waveform skew / 79
To display reference waveform information / 80
To save/recall reference waveform files to/from a USB storage device / 80
Analog channel or math waveforms can be saved to one of two reference
waveform locations in the oscilloscope. Then, a reference waveform can be
displayed and compared against other waveforms. One reference waveform can
be displayed at a time.
You can adjust the vertical scale and offset of reference waveforms. There is also a
skew adjustment for reference waveforms. Reference waveform scale, offset, and
skew information can optionally be included on the oscilloscope display.
Analog channel, math, or reference waveforms can be saved to a reference
waveform file on a USB storage device. You can recall a reference waveform file
from a USB storage device into one of the reference waveform locations.
To save a waveform to a reference waveform location
1 Press the [Analyze] key.
2 Press Features; then, select R1 or R2.
3 Press Features again (or push the Entry knob) to turn on the reference waveform.
4 Press Save/Clear.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight DSOX1202A and is the answer not in the manual?

Keysight DSOX1202A Specifications

General IconGeneral
BrandKeysight
ModelDSOX1202A
CategoryTest Equipment
LanguageEnglish

Related product manuals