EasyManua.ls Logo

Keysight DSOX1202A - I2 C Triggering and Serial Decode

Keysight DSOX1202A
330 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
26 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
1 Getting Started
1 Input the Probe Comp signal (see "Input a Waveform" on page 23).
2 Press [Default Setup] to recall the default oscilloscope setup (see "Recall the
Default Oscilloscope Setup" on page 24).
3 Press [Auto Scale] to automatically configure the oscilloscope for the Probe
Comp signal (see "Use Autoscale" on page 24).
4 Press the channel key to which the probe is connected ([1], [2], etc.).
5 In the Channel menu, press Probe.
6 In the Channel Probe menu, press Probe Check; then, follow the instructions
on-screen.
If necessary, use a nonmetallic tool (supplied with the probe) to adjust the
trimmer capacitor on the probe for the flattest pulse possible.
On some probes (like the N2140/42A probes), the trimmer capacitor is located
on the probe BNC connector. On other probes (like the N2862/63/90 probes),
the trimmer capacitor is a yellow adjustment on the probe tip.
7 Connect probes to all other oscilloscope channels.
8 Repeat the procedure for each channel.
Learn the Front Panel Controls and Connectors
On the front panel, key refers to any key (button) you can press.
NOTE
If your probe has a configurable attenuation setting (like the N2140/42A probes do), the 10:1
setting must be used for probe compensation.
Perfectly compensated
Over compensated
Under compensated

Table of Contents

Related product manuals