232 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
22 Utility Settings
 
2 In the Utility menu, press I/O.
3 In the I/O menu, press Configure.
• LAN — When connected to a LAN, you can use the LAN Settings and LAN Reset 
softkeys to configure the LAN interface. See "Setting up the Oscilloscope's 
LAN Connection" on page 232.
• USB — You can enable Compatibility Mode if you are experiencing connectivity 
issues when using the oscilloscope as a USB device. This mode provides 
maximum compatibility with USB host controllers at the cost of I/O 
performance.
When an I/O interface is present, remote control over that interface is always 
enabled. Also, the oscilloscope can be controlled via multiple I/O interfaces (for 
example, USB and LAN) at the same time.
See Also • Chapter 23, “Web Interface,” starting on page 249 (when the oscilloscope is 
connected to a LAN).
• The oscilloscope's Programmer's Guide.
Setting up the Oscilloscope's LAN Connection
You can place the oscilloscope on the network and set up its LAN connection. 
Once that is done, you can use the oscilloscope's web interface or remotely 
control the oscilloscope via the LAN interface.
The oscilloscope supports methods for automated LAN configuration or manual 
LAN configuration (see "To establish a LAN connection" on page 233). It is also 
possible to set up a point-to-point LAN connection between a PC and the 
oscilloscope (see "Stand-alone (Point-to-Point) Connection to a PC" on 
page 234).
Once the oscilloscope is set up on the network, you can use the oscilloscope's 
web page to view or change its network configuration and access additional 
settings (like the network password). See Chapter 23, “Web Interface,” starting on 
page 249.
Any time you modify the oscilloscope's hostname, it breaks the connection between the 
oscilloscope and the LAN. You need to re-establish communication to the oscilloscope using 
the new hostname.