52 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
3 Vertical Controls
2 In the Channel menu, press the Probe softkey to display the Channel Probe
menu.
The Channel Probe menu lets you select additional probe
parameters such as attenuation factor and units of
measurement for the connected probe.
The Probe Check softkey guides you through the process of
compensating passive probes (such as the N2140A, N2142A,
N2862A/B, N2863A/B, N2889A, N2890A, 10073C, 10074C,
or 1165A probes).
See Also • "To specify the channel units" on page 52
• "To specify the probe attenuation" on page 52
• "To specify the probe skew" on page 53
To specify the channel units
1 Press the probe's associated channel key.
2 In the Channel menu, press Probe.
3 In the Channel Probe menu, press Units; then, select:
• Volts — for a voltage probe.
• Amps — for a current probe.
Channel sensitivity, trigger level, measurement results, and math functions will
reflect the measurement units you have selected.
To specify the probe attenuation
The probe attenuation factor must be set properly for accurate measurement
results.