Waveform Generator 19
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 209
4 In the Wave Gen Mod menu:
Press the Modulation softkey to enable or disable
modulated waveform generator output.
You can enable modulation for all waveform generator
function types except pulse, DC, and noise.
Press the Type softkey and turn the Entry knob to select
the modulation type:
Amplitude Modulation (AM) — the amplitude of the
original carrier signal is modified according to the
amplitude of the modulating signal. See "To set up
Amplitude Modulation (AM)" on page 209.
Frequency Modulation (FM) — the frequency of the original
carrier signal is modified according to the amplitude of
the modulating signal. See "To set up Frequency
Modulation (FM)" on page 210.
Frequency-Shift Keying Modulation (FSK) — the output
frequency "shifts" between the original carrier frequency and a "hop
frequency" at the specified FSK rate. The FSK rate specifies a digital
square wave modulating signal. See "To set up Frequency-Shift Keying
Modulation (FSK)" on page 212.
To set up Amplitude Modulation (AM)
In the Wave Gen Mod menu (under [Wave Gen] > Settings > Modulation):
1 Press the Type softkey and turn the Entry knob to select Amplitude Modulation
(AM).
2 Press the Waveform softkey and turn the Entry knob to select the shape of the
modulating signal:
Sine
Square
Ramp
When the Ramp shape is selected, a Symmetry softkey appears so that you can
specify the amount of time per cycle that the ramp waveform is rising.
3 Press the AM Freq softkey and turn the Entry knob to specify the frequency of
the modulating signal.