224 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
20 Save/Recall (Setups, Screens, Data)
 
To recall reference waveform files from a USB storage device
1 Press the [Save/Recall] key.
2 In the Save/Recall menu, press the Recall softkey.
3 In the Recall menu, press the Recall softkey and turn the Entry knob to select 
Reference Waveform data (*.h5) .
4 Press the To Ref: softkey and turn the Entry knob to select the desired reference 
waveform location.
5 Press the softkey in the second position and use the Entry knob to navigate to 
the file to recall. See "To navigate storage locations" on page 221.
6 Press the Press to Recall softkey.
A message indicating whether the recall was successful is displayed.
7 If you would like to clear the display of everything except the reference 
waveform, press Clear Display.
Recalling Default Setups
1 Press the [Save/Recall] key.
2 In the Save/Recall menu, press Default/Erase.
3 In the Default menu, press one of these softkeys:
• Default Setup— recalls the oscilloscope's default setup. This is the same as 
pressing the front panel [Default Setup] key. See "Recall the Default 
Oscilloscope Setup" on page 24.
Some user settings are not changed when recalling the default setup.
• Factory Default— recalls the oscilloscope's factory default settings.
You must confirm the recall because there are no user settings that are left 
unchanged.
Performing a Secure Erase
1 Press the [Save/Recall] key.
2 In the Save/Recall menu, press Default/Erase.