224 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
20 Save/Recall (Setups, Screens, Data)
To recall reference waveform files from a USB storage device
1 Press the [Save/Recall] key.
2 In the Save/Recall menu, press the Recall softkey.
3 In the Recall menu, press the Recall softkey and turn the Entry knob to select
Reference Waveform data (*.h5) .
4 Press the To Ref: softkey and turn the Entry knob to select the desired reference
waveform location.
5 Press the softkey in the second position and use the Entry knob to navigate to
the file to recall. See "To navigate storage locations" on page 221.
6 Press the Press to Recall softkey.
A message indicating whether the recall was successful is displayed.
7 If you would like to clear the display of everything except the reference
waveform, press Clear Display.
Recalling Default Setups
1 Press the [Save/Recall] key.
2 In the Save/Recall menu, press Default/Erase.
3 In the Default menu, press one of these softkeys:
Default Setup— recalls the oscilloscope's default setup. This is the same as
pressing the front panel [Default Setup] key. See "Recall the Default
Oscilloscope Setup" on page 24.
Some user settings are not changed when recalling the default setup.
Factory Default— recalls the oscilloscope's factory default settings.
You must confirm the recall because there are no user settings that are left
unchanged.
Performing a Secure Erase
1 Press the [Save/Recall] key.
2 In the Save/Recall menu, press Default/Erase.