Horizontal Controls 2
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 45
 
The area of the normal display that is expanded is outlined with a box and the rest 
of the normal display is ghosted. The box shows the portion of the normal sweep 
that is expanded in the lower half.
To change the time/div for the Zoom window, turn the horizontal scale (sweep 
speed) knob. As you turn the knob, the zoomed window time/div is highlighted in 
the status line above the waveform display area. The Horizontal scale (sweep 
speed) knob controls the size of the box.
The Horizontal position (delay time) knob sets the left-to-right position of the 
zoom window. The delay value, which is the time displayed relative to the trigger 
point) is momentarily displayed in the upper-right portion of the display when the 
delay time ( ) knob is turned.
Negative delay values indicate you're looking at a portion of the waveform before 
the trigger event, and positive values indicate you're looking at the waveform after 
the trigger event.
To change the time/div of the normal window, turn off Zoom; then, turn the 
horizontal scale (sweep speed) knob.
For information about using zoom mode for measurements, refer to "To isolate a 
pulse for Top measurement" on page 161 and "To isolate an event for frequency 
measurement" on page 168.
To change the horizontal scale knob's coarse/fine adjustment 
setting
1 Push the horizontal scale knob to toggle between fine and coarse adjustment 
of the horizontal scale.
When Fine is enabled, turning the horizontal scale knob changes the time/div 
(displayed in the status line at the top of the display) in smaller increments. The 
time/div remains fully calibrated when Fine is on.
When Fine is turned off, the Horizontal scale knob changes the time/div setting in 
a 1-2-5 step sequence.