Measurements 15
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 167
Time Measurements
The following figure shows time measurement points.
The default lower, middle, and upper measurement thresholds are 10%, 50%, and
90% between Top and Base values. See "Measurement Thresholds" on page 175
for other percentage threshold and absolute value threshold settings.
"Period" on page 167
"Frequency" on page 168
"Counter" on page 169
"+ Width" on page 169
"– Width" on page 169
"Bit Rate" on page 169
"Duty Cycle" on page 170
"Rise Time" on page 170
"Fall Time" on page 170
"Delay" on page 170
"Phase" on page 171
"X at Min Y" on page 173
"X at Max Y" on page 173
Period
Period is the time period of the complete waveform cycle. The time is measured
between the middle threshold points of two consecutive, like-polarity edges. A
middle threshold crossing must also travel through the lower and upper threshold
levels which eliminates runt pulses. The X cursors show what portion of the
waveform is being measured. The Y cursor shows the middle threshold point.
Rise Time
+ Width
Fall Time
- Width
Period
Middle
Upper
Thresholds
Lower