SPI Triggering
To set up the oscilloscope to capture SPI signals, see "Setup for SPI Signals" on
page 301.
After the oscilloscope has been set up to capture SPI signals, you can then trigger
on a data pattern that occurs at the start of a frame. The serial data string can be
specified to be from 4 to 32 bits long.
When you select the SPI trigger type and Display Info is enabled, a graphic is
displayed showing the current state of the frame signal, clock slope, number of
data bits, and data bit values.
1 Press [Trigger].
2 In the Trigger menu, press the Trigger Type softkey.
3 Press the Trigger Type softkey; then, turn the Entry knob to select the serial slot
(Serial 1) on which the SPI signals are being decoded.