312 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
28 SPI Triggering and Serial Decode
 
Interpreting SPI Lister Data
In addition to the standard Time column, the SPI Lister contains these columns:
• Data — data bytes (MOSI and MISO).
Aliased data is highlighted in pink. When this happens, decrease the horizontal 
time/div setting and run again.