Triggers 11
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 111
8 Press the Auto Setup softkey to automatically set up the oscilloscope for the
selected Source and Standard:
Source channel vertical scaling is set to 140 mV/div.
Source channel offset is set to 245 mV.
Source channel is turned on.
Trigger type is set to Video.
Video trigger mode is set to All Lines.
Display Grid type is set to IRE (when Standard is NTSC) or mV (see "To select
the grid type" on page 86).
Horizontal time/division is set to 10 µs/div for NTSC/PAL/SECAM standards.
Horizontal delay is set so that trigger is at first horizontal division from the
left.
You can also press [Analyze]> Features and then select Video to quickly access the
video triggering automatic set up and display options.
9 Press the Mode softkey to select the portion of the video signal that you would
like to trigger on.
The Video trigger modes available are:
Field1 and Field2 — Trigger on the rising edge of the first serration pulse of
field 1 or field 2 (interlaced standards only).
All Fields — Trigger on the rising edge of the first pulse in the vertical sync
interval.
All Lines — Trigger on all horizontal sync pulses.
Line: Field1 and Line:Field2 — Trigger on the selected line # in field 1 or field 2
(interlaced standards only).
Line: Alternate — Alternately trigger on the selected line # in field 1 and field 2
(NTSC, PAL, PAL-M, and SECAM only).
Standard Type Sync Pulse
NTSC Interlaced Bi-level
PAL Interlaced Bi-level
PAL-M Interlaced Bi-level
SECAM Interlaced Bi-level