Waveform Generator 19
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 211
 
When the Ramp shape is selected, a Symmetry softkey appears so that you can 
specify the amount of time per cycle that the ramp waveform is rising.
3 Press the FM Freq softkey and turn the Entry knob to specify the frequency of the 
modulating signal.
4 Press the FM Dev softkey and turn the Entry knob to specify the frequency 
deviation from the original carrier signal frequency.
When the modulating signal is at its maximum amplitude, the output frequency 
is the carrier signal frequency plus the deviation amount, and when the 
modulating signal is at its minimum amplitude, the output frequency is the 
carrier signal frequency minus the deviation amount.
The frequency deviation cannot be greater than the original carrier signal 
frequency.
Also, the sum of the original carrier signal frequency and the frequency 
deviation must be less than or equal to the maximum frequency for the selected 
waveform generator function plus 100 kHz.
The following screen shows an FM modulation of a 100 kHz sine wave carrier 
signal.