294 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
27 LIN Triggering and Serial Decode
 
6 Press the Signals softkey to open the LIN Signals menu.
7 Press the Source softkey to select the channel connected to 
the LIN signal line.
The label for the LIN source channel is automatically set.
8 Press the Threshold softkey; then, turn the Entry knob to set 
the LIN signal threshold voltage level to the middle of the LIN 
signal.
The threshold voltage level is used in decoding, and it will 
become the trigger level when the trigger type is set to the 
selected serial decode slot.
9 Press the Baud Rate softkey to open the LIN Baud Rate menu.
10 Press the Baud softkey; then, turn the Entry knob to select the 
baud rate that matches match your LIN bus signal.
The default baud rate is 19.2 kb/s.
If none of the pre-defined selections match your LIN bus 
signal, select User Defined; then, press the User Baud softkey and turn the Entry 
knob to enter the baud rate.
You can set the LIN baud rate from 2.4 kb/s to 625 kb/s in increments of 
100 b/s.
11 Press the  Back key to return to the LIN Signals menu.
12 Press the Sample Point softkey; then, turn the Entry knob to select the sample 
point at which the oscilloscope will sample the bit value.
13 Press the Standard softkey; then, turn the Entry knob to select the LIN standard 
you are measuring (LIN 1.3 or LIN 2.0).
60%
70%
80%
Sample Point
One Bit