294 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
27 LIN Triggering and Serial Decode
6 Press the Signals softkey to open the LIN Signals menu.
7 Press the Source softkey to select the channel connected to
the LIN signal line.
The label for the LIN source channel is automatically set.
8 Press the Threshold softkey; then, turn the Entry knob to set
the LIN signal threshold voltage level to the middle of the LIN
signal.
The threshold voltage level is used in decoding, and it will
become the trigger level when the trigger type is set to the
selected serial decode slot.
9 Press the Baud Rate softkey to open the LIN Baud Rate menu.
10 Press the Baud softkey; then, turn the Entry knob to select the
baud rate that matches match your LIN bus signal.
The default baud rate is 19.2 kb/s.
If none of the pre-defined selections match your LIN bus
signal, select User Defined; then, press the User Baud softkey and turn the Entry
knob to enter the baud rate.
You can set the LIN baud rate from 2.4 kb/s to 625 kb/s in increments of
100 b/s.
11 Press the Back key to return to the LIN Signals menu.
12 Press the Sample Point softkey; then, turn the Entry knob to select the sample
point at which the oscilloscope will sample the bit value.
13 Press the Standard softkey; then, turn the Entry knob to select the LIN standard