298 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
27 LIN Triggering and Serial Decode
4 Press the Mode softkey; then, select LIN.
5 Choose whether to include the parity bits in the identifier
field.
a If you want to mask the upper two parity bits, ensure that
the box under the Show Parity softkey is not selected.
b If you want to include the parity bits in the identifier field,
ensure that the box under the Show Parity softkey is
selected.
6 If the oscilloscope is stopped, press the [Run/Stop] key to
acquire and decode data.
You can use the horizontal Zoom window for easier navigation of the decoded data.
See Also • "Interpreting LIN Decode" on page 299
• "Interpreting LIN Lister Data" on page 300
If the setup does not produce a stable trigger, the LIN signal may be slow enough that the
oscilloscope is auto-triggering. Press the [Trigger] key, then press the Mode softkey to set the
trigger mode from Auto to Normal.