310 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
28 SPI Triggering and Serial Decode
4 Press the Mode softkey; then, select SPI.
5 Press the Word Size softkey; then, turn the Entry knob to
select the number of bits in a word.
6 Press the Bit Order softkey; then turn the Entry knob to select
the bit order, most significant bit first (MSB) or least
significant bit first (LSB), used when displaying data in the
serial decode waveform.
7 If the oscilloscope is stopped, press the [Run/Stop] key to
acquire and decode data.
You can use the horizontal Zoom window for easier navigation of the acquired data.
See Also "Interpreting SPI Decode" on page 311
"Interpreting SPI Lister Data" on page 312
NOTE
If the setup does not produce a stable trigger, the SPI signal may be slow enough that the
oscilloscope is auto-triggering. Press the [Trigger] key, then press the Mode softkey to set the
trigger mode from Auto to Normal.