EasyManua.ls Logo

Keysight DSOX1202A - Page 319

Keysight DSOX1202A
330 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
UART/RS232 Triggering and Serial Decode 29
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 319
6 In the UART Settings menu, press the Base softkey to select the base (hex,
binary, or ASCII) in which decoded words are displayed.
When displaying words in ASCII, the 7-bit ASCII format is
used. Valid ASCII characters are between 0x00 and 0x7F.
To display in ASCII you must select at least 7 bits in the
Bus Configuration. If ASCII is selected and the data
exceeds 0x7F, the data is displayed in hex.
When #Bits is set to 9 in the UART Bus Config menu, the
9th (alert) bit is displayed directly to the left of the ASCII
value (which is derived from the lower 8 bits).
7 Optional: Press the Framing softkey and select a value. In the
decode display, the chosen value will be displayed in light
blue. However, if a parity error occurs the data will be
displayed in red.
8 If the oscilloscope is stopped, press the [Run/Stop] key to
acquire and decode data.
You can use the horizontal Zoom window for easier navigation of the acquired data.
See Also "Interpreting UART/RS232 Decode" on page 320
"UART/RS232 Totalizer" on page 321
"Interpreting UART/RS232 Lister Data" on page 322
NOTE
If the setup does not produce a stable trigger, the UART/RS232 signals may be slow enough
that the oscilloscope is auto-triggering. Press the [Trigger] key, then press the Mode softkey
to set the trigger mode from Auto to Normal.