X-Stream Operator’s Manual
WM-OM-E Rev I 245
This selection means, for example, that if you select All In the test will be False if even a
single waveform falls outside the mask.
6. From Show Markers, choose whether or not to have mask violations displayed.
7. If you are loading a pre-existing mask, touch the Load Mask tab, then the File button. You
can then enter the file name or browse to its location.
8. If you want to make a mask from your waveform, touch the Make Mask tab.
9. Touch inside the Ver Delta and Hor Delta fields and enter boundary values, using the
pop-up numeric keypad.
10. Touch the Browse button to create a file name and location for the mask if you want to
save it.
11. Touch the Gate tab, then enter values in the Start and Stop fields to constrain the mask to
a portion of the waveform. Or, you can simply touch and drag the Gate posts, which initially
are placed at the extreme left and right ends of the grid.