EasyManuals Logo

LeCroy WaveMaster 8600A User Manual

LeCroy WaveMaster 8600A
437 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #33 background imageLoading...
Page #33 background image
X-Stream Operator’s Manual
WM-OM-E Rev I 31
compatible probes; supports ProLink-SMA and ProLink-BNC adapters (ProLink is not available for
WavePro 7000A series)
Scale Factors: Automatically or manually selected depending on probe used
AP-1M Hi-Z Adapter: (not available for WavePro 7000A series) Bandwidth: 500 MHz; full-scale
range: +/-8 V; input protection: +/-150 V
Color Waveform Display
Type: Color 10.4-inch flat panel TFT LCD with high resolution touch screen
Resolution: SVGA; 800 x 600 pixels
Real Time Clock: Date, hours, minutes, and seconds displayed with waveform; SNTP support to
synchronize to precision internet clocks
Number of Traces: Maximum of eight traces; simultaneously displays channel, zoom, memory,
and math traces
Grid Styles: Single, Dual, Quad, Octal, XY, Single+XY, Dual+XY
Waveform Display Styles: Sample dots joined or dots only
Analog Persistence Display
Analog and Color-graded Persistence: Variable saturation levels; stores each trace's
persistence data in memory
Persistence Selections: Select analog, color, or 3-D
Trace Selection: Activate Analog Persistence on all or any combination of traces
Persistence Aging Time: From 500 ms to infinity
Sweeps Displayed: All accumulated or all accumulated with last trace highlighted
Zoom Expansion Traces
Display up to 4 Zoom and 4 Math/Zoom traces; 8 Math/Zoom traces available with XMAP (Master
Analysis Package) and XMATH (Advanced Math Package) options.
Rapid Signal Processing
Processor: Intel® Pentium 4 @ 2.53 GHz (or better) with MS Windows® XP Platform
Processor Memory: Up to 1 Gbyte (WaveMaster: up to 2 Gbytes with XXL memory option)
Internal Waveform Memory
Waveform: M1, M2, M3, M4 (Store full-length waveforms with 16 bits/data point.) Or save to any
number of files (limited only by data storage media).
Setup Storage
Front Panel and Instrument Status: Save to the internal hard drive, floppy drive, or to a USB
connected peripheral device.

Table of Contents