X-Stream Operator’s Manual
WM-OM-E Rev I 37
Warning
This is a Class A product. In a domestic environment this product may cause radio interference, in
which case the user may be required to take appropriate measures.
EN 61000-4-2/A2:2001* Electrostatic Discharge
(4 kV contact, 8 kV air, 4 kV vertical/horizontal coupling planes)
EN 61000-4-3/A1:2003* RF Radiated Electromagnetic Field
(3 V/m, 80-1000 MHz)
EN 61000-4-5/A1:2001* Electrical Fast Transient/Burst
(1 kV AC Mains, 0.5 kV I/O signal/control)
EN 61000-4-5:1995* Surges
(1 kV AC Mains, 0.5 kV I/O signal/control)
EN 61000-4-6/A1:2001* RF Conducted Electromagnetic Field
(1 kV / 0.5 kV common mode / differential mode - AC Mains)
Electromagnetic Immunity:
EN 61000-4-11:2004
†
Mains Dips and Interruptions
(1 cycle voltage dip, 100% short interruption)
* Meets Performance Criteria "B" limits during the disturbance, product undergoes a
temporary degradation or loss of function of performance which is self recoverable.
† Meets Performance Criteria "C" limits during the disturbance, product undergoes a
temporary degradation or loss of function of performance which requires operator
intervention or system reset.
Low Voltage Directive: EN 61010-1:2001
Safety requirements for electrical equipment for measurement,
control, and laboratory use.
The oscilloscope has been qualified to the following EN 61010-1
limits:
Installation Categories II (Mains Supply Connector) & I
(Measuring Terminals)
Pollution Degree 2 (Normally only dry non-conductive pollution
occurs. Occasionally a temporary conductivity caused by
condensation must be expected.)
Protection Class I (Provided with terminal for protective ground)
UL Standard: UL 3111-1
UL and cUL Certifications:
Canadian Standard: CSA-C22.2 No. 1010.1-92