EasyManuals Logo

LINET Sprint 200 User Manual

LINET Sprint 200
136 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #107 background imageLoading...
Page #107 background image
D9U001ES2-0101_05
107
17.11.11 Technical Specications
Parameter Value
i-Drive Power wheel diameter 21 cm
Max. fast forward speed (at ground, loaded) 4,43 km/h (±15%)
Max. forward speed (at ground, loaded) 2,16 km/h (±15%)
Max. reverse speed (at ground, loaded) 2,16 km/h (±15%)
Max. angle of ascent
17.11.12 Electrical specication
Parameter Value
Input Voltage, Frequency 230 V AC, 50/60 Hz
127 V AC, 50/60 Hz
120 V AC, 50/60 Hz
110 V AC, 50/60 Hz
100 V AC, 50/60 Hz
Batteries Voltage 12 V DC, Capacity: 9 Ah
Maximum Power Input 300 W
Fuse
Version 230 V
Version 127 V
Version 120 V
Version 110 V
Version 100 V
2 x T1,6A L 250V
2 x T3,15A L 250V
2 x T3,15A L 250V
2 x T3,15A L 250V
2 x T3,15A L 250V
17.11.13 i-Drive Power Maintenance
Periodical maintenance of the i-Drive Power must be done by qualied service technician or authorized service organization at least
once a year. To continue maintenance please see chapter Maintenance.
Service technician must check the following:
accumulator status and eventual replacement of the accumulator (after maximum of three years of duty)
shock absorber function
i-Drive Power wheel – replace if necessary
lifting mechanism – grease if necessary
cables, control elements – replace if necessary
i-Drive Power function

Table of Contents

Other manuals for LINET Sprint 200

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LINET Sprint 200 and is the answer not in the manual?

LINET Sprint 200 Specifications

General IconGeneral
BrandLINET
ModelSprint 200
CategoryMedical Equipment
LanguageEnglish

Related product manuals