D9U001ES2-0101_0586
16.9 Settings Menu
Operator is authorized to display scales veri cation, to check software and hardware versions, to set time and date and to set time
and date format in the Settings Menu.
To enter Veri cation Screen:
► Use ON/OFF Button or PAUSE Button to select the line
VERIFICATION.
► Press WEIGHT/CLEAR Button to con rm the selection.
To enter Versions Screen:
► Use ON/OFF Button or PAUSE Button to select the line
VERSIONS.
► Press WEIGHT/CLEAR Button to con rm the selection.
To enter Settings Menu:
► Press Bed Exit Alarm Monitoring Button, PAUSE Button and ON/OFF Button at the same time for 3s.
Settings Menu is opened only for 60s unless adjustment follows.
Fig. Settings Menu Fig. Veri cation Screen
Fig. Software and Hardware Versions Screen
Parameter Meaning
BSCD version of the LW20 scales
system
LCDS version of the control panel
with display
CALIBRATION COUNTER number of changes of the
legally relevant parameters
(e.g. calibrations)
G LOC local gravitational constant
G CAL calibrated gravitational
constant