24.3 Safety Technical Checks
24 Maintenance (Sprint 200 with scales or with i-Drive Power)
24.1 Regular maintenance
Check regularly movable parts for wear.
Perform regularly visual check of the product (with delivery note if necessary).
Ask service department of the manufacturer for addition of the original spare parts if some product parts are missing.
Ask service department of the manufacturer for replacement of any damaged product parts by the original spare parts.
Check that the batteries are working properly. Disconnect the stretcher from the mains power to check signalisation of
battery indicator according to the instructions for use.
Have the batteries replaced if they are not working properly.
Check regularly that all accessories are working properly.
Replace damaged accessories immediately.
24.2 Spare Parts
The serial label is located on the frame of the mattress support platform. The serial label contains information for claims and orde-
ring replacement parts.
Information about spare parts is available from:
Manufacturer´s customer service
Sales department
WARNING!
Risk of injury when working on the stretcher!
Ensure that the stretcher is disconnected from the mains power prior to installation, putting into service, main-
tenance and deinstallation.
Ensure that the castors are locked prior to installation, putting into service, maintenance and deinstallation.
No part of the Sprint 200 ME equipment shall be serviced or maintained while in use with a patient.
WARNING!
Risk of injury due to defective stretcher!
Have a defective stretcher repaired immediately.
If the defect cannot be repaired, do not use the stretcher.
CAUTION!
Material damage due to incorrect maintenance!
Ensure that maintenance is performed exclusively by manufacturer´s customer service or by authorised servi-
ce personnel certied by the manufacturer.
If the defect cannot be repaired, do not use the stretcher.
WARNING!
Risk of injury due to incorrect safety technical checks!
Ensure that safety technical checks are performed exclusively by manufacturer´s customer service or by
authorised service personnel certied by the manufacturer.
Ensure that the safety technical checks are recorded in the service and maintenance log.
LINET ® recommends attaching the maintenance plaque to the stretcher.
Safety technical check of the stretcher must be performed at least once every 12 months.
The procedure for performing the safety technical check is stipulated in EN 62353:2014.
NOTE On request, the manufacturer will provide service documentation (e.g. circuit diagrams, component part lists, descriptions,
calibration instructions etc.) for service personnel for the repair of ME equipment designated by the manufacturer as repairable by
service personnel.

Table of Contents

Other manuals for LINET Sprint 200

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LINET Sprint 200 and is the answer not in the manual?

LINET Sprint 200 Specifications

General IconGeneral
BrandLINET
ModelSprint 200
CategoryMedical Equipment
LanguageEnglish

Related product manuals