Error/Fault Cause Solution
Faulty Mattress Support Platform Height
Adjustment
Obstacle on the undercarriage cover. Remove the obstacle.
Obstacle under the pedals. Remove the obstacle.
Faulty pedal. Notify the manufacturer´s service
department.
Lowering Backrest from the upright position
not possible
Obstacle under the Backrest or in the drive
mechanism.
Remove the obstacle.
Backrest Release Handle is defective. Notify the manufacturer´s service
department.
Adjusting Siderails not possible Obstacle in the Siderail Release Mechanism. Remove the obstacle.
Siderail Release Mechanism is defective. Notify the manufacturer´s service
department.
Unlocking Siderail not possible Incorrect Use. Do not push the siderail towards
stretcher head end when unloc-
king the siderail by the yellow
Siderail Release Levers!
Ensure no patient and no mattre-
ss push the siderail towards stret-
cher head end when unlocking
the siderail by the yellow Siderail
Release Levers!
Push the siderail slightly towards
stretcher foot end to facilitate ma-
nipulation with the yellow Siderail
Release Levers if needed.
Faulty brakes Obstacle blocking brakes mechanically. Remove the obstacle.
The brake mechanism is defective. Notify the manufacturer´s service
department.
Bed Exit Alarm Monitoring cannot be acti-
vated.
Sprint 200 with scales is disconnected from
the mains.
Connect Sprint 200 with scales to
the mains.
Scales and Bed Exit Alarm Monitoring Cont-
rol Panel indicates no battery.
Incorrectly inserted batteries. Replace the 4 batteries correctly.
Negative weight value is displayed on the
display.
No Taring was performed. Zero the scales.
Stabilized Scales Icon is ashing all the
time.
Scales are not stabilized. Disconnect Sprint 200 with scales
from the mains and do not touch
the Sprint 200 with scales.
Indicator of the stretcher disconnected from
the mains is lit all the time.
Power Supply Cord is disconneted from the
mains or from the Sprint 200 with scales.
Ensure the Sprint 200 with scales
is correctly connected to the
mains power.
BED EXIT WAITING is shown on the dis-
play.
Insu󰀩cient weight on the mattress support
platform of Sprint 200 with scales is detected.
Place patient on the mattress
support platform before activation
of the Bed Exit Alarm Monitoring.
Minimum patient weight for Bed
Exit Alarm Monitoring is 35kg.