7.5 Electromagnetic Compatibility (only Sprint 200 with scales or with i-
-Drive Power)
Stretcher is intended for hospitals except for near active HF surgical equipment and the RF shielded room of a medical
system for magnetic resonance imaging, where the intensity of EM disturbances is high.
Stretcher has dened no essential performance.
7.5.1 Manufacturer instructions - electromagnetic emissions
Emission Test Compliance
RF emissions
CISPR 11
Group 1
RF emissions
CISPR 11
Class A
Harmonic emissions
IEC 61000-3-2
Class A
Voltage uctuations / icker emissions
IEC 61000-3-3
Complies
NOTE The EMISSIONS characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR 11
class A). If it is used in a residential environment (for which CISPR 11 class B is normally required) this equipment might not o󰀨er
adequate protection to radio-frequency communication services. The user might need to take mitigation measures, such as reloca-
ting or re-orienting the equipment.
WARNING!
It is recommended to avoid the use of this device next to or in block with other device, because it could lead
to improper operation. If such use is needed, this device and the other equipment should be under surveillan-
ce to verify proper operation.
WARNING!
Use of the accessories, converters and other cables, than specied and provided by manufacturer of this 
stretcher could lead to increase of electromagnetic emission or lower the electromagnetic immunity of this
stretcher and lead to improper operation.
WARNING!
Mobile RF communication device (including end use devices like antenna cables and external antenna) should
not be used closer than 30 cm (12 inches) from any part of this stetcher Sprint 200 with scales or with i-Drive
Power, including cables specied by manufacturer. Otherwise this could lead to deterioration of functionality 
of this stretcher.
WARNING!
Do not overload the stretcher (SWL) and consider chapter 24 Maintenance in order to maintain the basic safe-
ty with regard to electromagnetic disturbances for the expected service life.
List of used cables:
Mains cable, maximum length 5 m

Table of Contents

Other manuals for LINET Sprint 200

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LINET Sprint 200 and is the answer not in the manual?

LINET Sprint 200 Specifications

General IconGeneral
BrandLINET
ModelSprint 200
CategoryMedical Equipment
LanguageEnglish

Related product manuals