D9U001ES2-0101_0578
15.4 Stretcher Overload
If load on the stretcher is over 330 kg:
► Warning popu-up window is shown on the display.
15.5 Stretcher Underload
If the stretcher is underloaded:
► Display shows the „LOW“.
15.6 Weighing in tilt
It is possible to use the scales in any position of the mattress support platform of the Sprint 200 with scales if its undercarriage is
situated on the horizontal oor.
15.7 Calibration of the Zero
Calibration of the Zero is only possible in a range of ±6,4kg from factory zero. Calibration of the Zero is used to reset weight on the
display and set up user zero, which sets the maximum weight range of the weighing system. Calibration of the Zero must be done
with an empty, unloaded stretcher, without the mattress and accessories. Calibration of the Zero is done after installation, weight
veri cation or servicing.
To calibrate the Zero:
► Position the stretcher about 20 cm above the lowest position and set the mattress support platform to the horizontal positi-
on. Ensure that nothing touches the stretcher except you.
► Press and hold ZERO/T Button until weight value starts to ash. Release ZERO/T Button.
► Press ZERO/T Button to con rm Calibration of the Zero.
“0” is shown on the display and a long acoustic signal con rms Calibration of the Zero.
To cancel the Calibration of the Zero:
► Press WEIGHT/CLEAR Button while Calibration of the Zero.
Fig. Sprint 200 with scales is
overloaded (pop-up)