D9U001ES2-0101_05128
20.1.3 Complete Cleaning and Disinfection
Clean the following stretcher parts:
■ All control elements for adjusting the stretcher
■ All handles
■ Head End and Foot End
■ Siderails (in highest position)
■ Freely accessible mattress surface
■ Accessory rails
■ All Mattress Support Platform covers
■ Plastic undercarriage covers
■ Column covers
■ Mattress on all sides
■ Freely accessible metal parts of Mattress Support Platform
■ Infusion stand sleeve tting
■ Bumpers
■ Castors
■ Brakes
21 Troubleshooting (Sprint 200 without scales and without i-
-Drive Power)
Error/Fault Cause Solution
Faulty Mattress Support Platform Height
Adjustment
Obstacle on the undercarriage cover. Remove the obstacle.
Obstacle under the pedals. Remove the obstacle.
Faulty pedal. Notify the manufacturer´s service
department.
Lowering Backrest from the upright position
not possible
Obstacle under the Backrest or in the drive
mechanism.
Remove the obstacle
Backrest Release Handle is defective. Notify the manufacturer´s service
department.
Adjusting Siderails not possible Obstacle in the Siderail Release Mechanism. Remove the obstacle.
Siderail Release Mechanism is defective. Notify the manufacturer´s service
department.
Unlocking Siderail not possible Incorrect Use. Do not push the siderail towards
stretcher head end when unloc-
king the siderail by the yellow
Siderail Release Levers!
Ensure no patient and no mattre-
ss push the siderail towards stret-
cher head end when unlocking
the siderail by the yellow Siderail
Release Levers!
Push the siderail slightly towards
stretcher foot end to facilitate ma-
nipulation with the yellow Siderail
Release Levers if needed.
Faulty brakes Obstacle blocking brakes mechanically. Remove the obstacle.
The brake mechanism is defective. Notify the manufacturer´s service
department.