EasyManua.ls Logo

LINET Sprint 200

LINET Sprint 200
136 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
D9U001ES2-0101_0532
7.3 Electrical Specications (only Sprint 200 with scales or with i-Drive 
Power)
Parameter Value
Input Voltage, Frequency (Sprint 200 only with scales) 100 — 240V AC, 50/60 Hz
Input Voltage, Frequency (Sprint 200 only with i-Drive Power) 100V AC, 50/60 Hz
110V AC, 50/60 Hz
120V AC, 50/60 Hz
127V AC, 50/60 Hz
230V AC, 50/60 Hz
Input Voltage, Frequency (Sprint 200 with scales and with i-Drive
Power)
100V AC, 50/60 Hz
110V AC, 50/60 Hz
120V AC, 50/60 Hz
127V AC, 50/60 Hz
230V AC, 50/60 Hz
Maximum Power Input (Sprint 200 only with scales) 24 VA
Maximum Power Input (Sprint 200 only with i-Drive Power) 400 VA
Maximum Power Input (Sprint 200 with scales and with i-Drive
Power)
400 VA
Ingress Protection according to EN 60529
(Sprint 200 with scales or with i-Drive Power)
IPX4
Batteries of the scales system 4 x AA LR6 1,5V (6V DC)
Batteries of the i-Drive Power system 3 x 12V 9Ah VRLA
Fuses in the Battery Box for scales system T1A
Fuses in the Power supply for scales system 2 x T1A L 250V
Fuses in the i-Drive Power system
Version 100V AC
Version 110V AC
Version 120V AC
Version 127V AC
Version 230V AC
2 x T3,15A L 250V
2 x T3,15A L 250V
2 x T3,15A L 250V
2 x T3,15A L 250V
2 x T1,6A L 250V
Electric Protection Class (Sprint 200 with scales or with i-Drive
Power)
Class I