16.10 Basic technical parameters of the Bed Exit Alarm Monitoring
Parameter Value Unit
Minimum Load for Bed Exit Alarm Monitoring activation Conguration Parameter (CP)
(5 000 – 500 000)
g
Dimensions of the monitored Inner Zone CP (100 – 1 000 × 100 – 1 000) mm × mm
Minimum decrease of weight to alarm activation CP (1 – 50) %
Automatic alarm deactivation after patient’s return to
stretcher
CP (0 – (2
32
− 1)) ms
Period of interrupted monitoring (PAUSE) after the press
of PAUSE button
CP (60 – 3600) s
Period of patient’s being outside the stretcher to the auto-
matic end of PAUSE after patient’s return to the stretcher
≥ 5 s
System reaction to the decrease of weight or to the patien-
t’s leaving the monitored zone
≤ 1,5 s
Alarm activation during scales system failure and activa-
ted monitoring
≤ 3 s
Alarm activation during scales system failure ascertained
immediately after (re)start if the monitoring was activated
before switching o󰀨
≤ 1,5 s