Cleaning and Disinfection Maintenance
7 - 12 046-004667-00 A7™ Operating Instructions
4. Remove the bellows adapter ring from inside the bellows (see FIGURE 7-8). Note the
orientation of the bellows adapter ring as it is being removed to ensure that it is properly
inserted during reassembly. (If the ring contains grooves, the ring should be oriented so that
the grooves are facing downward in the final reassembly.)
FIGURE 7-8 Removing the Bellows Adapter Ring
5. Remove the bellows dome O-ring (see FIGURE 7-9).
FIGURE 7-9 Removing the Bellows Dome O-ring
6. Cleaning
a. To prevent damage, wash each component gently using a recommended cleaning
agent (see TABLE 7-2 on page 7-8). Ensure that all bellows surfaces are cleaned. Do not
autoclave the bellows.
b. Rinse with clean, hot water, and allow to dry.
NOTE: Dry the bellows by allowing it to hang so that it is fully expanded. This
will facilitate thorough drying and prevent it from sticking to itself.
CAUTION: Do not autoclave the following components: Paw gauge, flow sensor,
and bellows. These components cannot withstand immersion or the
heat and pressure of autoclaving.
Grooves on adapter ring face
downward.

Table of Contents

Other manuals for Mindray A7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A7 and is the answer not in the manual?

Mindray A7 Specifications

General IconGeneral
TypeAnesthesia Workstation
BrandMindray
ModelA7
Display15-inch color touchscreen
Oxygen FlushYes
Gas InletsO2, N2O, Air
Ventilation ModesSIMV, PSV
Breathing SystemIntegrated
Gas MonitoringO2, CO2, N2O
Monitoring ParametersFiO2
CO2 MonitoringMainstream or Sidestream
Power Supply100-240 V, 50/60 Hz

Related product manuals